Sensorex S660CD In-line pH Sensor for 600 Series Assemblies
Sensorex S660CD In-line pH Sensor for 600 Series Assemblies
The Sensorex S660CD is a flat-surface, self-cleaning CPVC pH electrode for in-line mounting in a 3/4″ pipe tee. The S661CD and S662CD are the 1″ and 2″ versions of the same family. It drops into Sensorex’s reusable FC66P gland or FC66-series flow cell. Sensorex recommends this in-line mounting for replacing existing threaded electrodes. Maintenance stays simple: no tools, a quick-disconnect BNC, and the gland stays in the line. The sealed Ag/AgCl double-junction reference needs no filling solution. The flat glass resists coating when the electrode sits in turbulent flow.
Low-ionic (-LC) and acidic-fluoride (-HF) variants and ORP versions are offered on the same platform. Automatic temperature compensation comes from the FC66TC, FC67TC or FC68TC flow-cell-and-cable kits. Jadler Industries stocks the S660 family and matching glands and flow cells in Calgary. They serve water treatment, chemical dosing and process pH loops across Western Canada.
- Features & Benefits
- Applications
- Product manuals
- Product specs
- Quick-disconnect design saves time: installs in seconds, no tools
- Flat-surface glass resists coating in turbulent flow
- Ideal for replacing existing threaded in-line electrodes
- Reusable gland — electrodes, glands and flow cells for 3/4″, 1″ and 2″ lines
- Sealed double-junction reference, no refilling
- Available with or without ATC through the FC66TC / FC67TC / FC68TC kits
- Standard, low-ionic and HF-resistant models, plus ORP versions
- In-line piping where flow can be shut off for electrode maintenance
- Self-cleaning installations in turbulent flow
- Replacing existing threaded in-line electrodes
- Low-ionic water (-LC version)
- Streams containing acidic fluorides (-HF version)
| pH range | 0–14 pH (0–12 pH without Na+ error) |
|---|---|
| Body / junction | CPVC / HDPE |
| Temperature / pressure | 0–75 °C at 0–100 psig; 0–80 °C at 0–85 psig |
| Reference | Ag/AgCl, sealed double junction |
| Line size | 3/4″ (S660CD); 1″ (S661CD); 2″ (S662CD) |
| Electrode dimensions (S660CD) | Dim. A 2.50″, Dim. B 4.25″, Dim. C 0.75″ (see spec-sheet drawing) |
| Mounting gland | FC66P: 20% glass-filled polypropylene, 3/4″ NPT, 1 1/16-20 UNF compression nut |
| Cable (sold separately) | S653 cap cable, RG174 coaxial with noise-reduction layer; ATC kits FC66TC / FC67TC / FC68TC include the cable |
| Temperature compensation | FC66TC / FC67TC / FC68TC flow-cell kits |
| Variants | -LC low ionic, -HF acidic fluorides, S660CD-ORP |
| Compliance | CE (listed in the Sensorex CE technical file); RoHS 3 |